When using a disposable human arterial blood sample collector, should we consider the interference of additives on the sample analysis results?
27 05 2024
Yes, the interference of additives on sample analysis results should be considered. The currently used anticoagulant is heparin. Known interferences of heparin with results include dilution effects, sodium heparin causing elevated sodium ion results, lithium heparin binding to calcium ions resulting in lower calcium ion results, and effects on pH levels. For products claimed to be suitable for testing corresponding items in blood (such as pH, potassium, sodium, calcium, chloride, electrolytes, hemoglobin, hematocrit, glucose, and lactate), research data must be submitted to demonstrate that the product does not interfere with the testing of these blood components, or that any interference is within acceptable limits.