What are the requirements for the cooling spray function when using a laser therapy device for hair removal? Should the cooling spray function be considered during performance, safety, and electromagnetic compatibility testing?
07 05 2024
Applicants should define the performance indicators related to the cooling spray function based on its mode of use, functional design, and whether it can be switched on or off. These indicators might include duration, interval time, and cooling effect. Applicants must also provide the basis for determining these indicators and the testing methods. During performance and safety testing, the cooling spray function should be considered and tested in conjunction with the cooling spray. For electromagnetic compatibility testing, the impact of activating the cooling spray function on the product should be considered, and testing should be conducted under the most adverse conditions.