What issues should be considered in the registration documents for large imaging equipment that provides a third-party physiological gating signal interface but does not include the gating device?
21 03 2024
For large imaging equipment (such as CT, MR, PET/CT, etc.) that provides a third-party physiological gating interface, such as a respiratory gating interface or an ECG gating interface, it should be clearly stated in the summary documents the requirements for third-party gating devices that can be used in conjunction, such as interface type (including connection methods, data protocols, etc.), compliance standards, etc. If it is a proprietary interface, the manufacturer and product model of compatible devices should also be specified. Research documents should provide verification and confirmation materials for tests conducted with third-party devices. The product technical requirements should specify the interface type (including connection methods, data protocols, etc.) and compliance standards (if applicable), and technical indicators related to gating should be tested.